Qualità Casse per la riparazione di dispositivi medici di GE RIC5-9-D Fabbrica

Casse per la riparazione di dispositivi medici di GE RIC5-9-D

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...

Qualità OLYMPUS CF-HQ190L Casse per la riparazione di dispositivi medici Casse per la riparazione di attrezzi Fabbrica

OLYMPUS CF-HQ190L Casse per la riparazione di dispositivi medici Casse per la riparazione di attrezzi

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...

Qualità Trasduttore convesso GE 4D3C-L 4D per riparazione Fabbrica

Trasduttore convesso GE 4D3C-L 4D per riparazione

GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome ...

Qualità Sonda addominale GE 4C-RS da riparare Fabbrica

Sonda addominale GE 4C-RS da riparare

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...

Qualità S5-2 Sonda cardiaca da riparare Fabbrica

S5-2 Sonda cardiaca da riparare

S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain ...

Qualità Toracoscopio Olympus LTF-160 per riparazione Fabbrica

Toracoscopio Olympus LTF-160 per riparazione

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...

Qualità OLYMPUS CF-H180AL Colonscopio elettronico per la riparazione Fabbrica

OLYMPUS CF-H180AL Colonscopio elettronico per la riparazione

OLYMPUS CF-H180AL Electronic Colonoscope For Repair ► Brief Description Inspection Results: 1. The image is dark, and the CCD is aged. 2. One branch of the light beam is disconnected. 3. The insertion tube is aged and wrinkled. 4. The lower angle is malfunctioning, the other angles are tight, and ...

Qualità SonoScape EG-500 Gastroscopia elettronica per riparazione Fabbrica

SonoScape EG-500 Gastroscopia elettronica per riparazione

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The ...

Qualità Olympus Flexible Endoscope Air & Water Tube A/W Transparent Tube Fabbrica

Olympus Flexible Endoscope Air & Water Tube A/W Transparent Tube

Olympus/ Pentax/ Fuji A/W Tube ► Brief Description The "air/water channel" is a critical part of a flexible endoscope's function. The tube you are referring to is most commonly known as the Air/Water Valve Tube or A/W Tube . It's a small, transparent, and highly flexible tube that connects the air ...