logo

Casse per la riparazione di dispositivi medici

Qualità Olympus CF-H190L Colonoscopio per la riparazione Fabbrica

Olympus CF-H190L Colonoscopio per la riparazione

Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; ...
Qualità Samsung Medison P2-4AH Sonda cardiaca per la riparazione Fabbrica

Samsung Medison P2-4AH Sonda cardiaca per la riparazione

Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2...
Qualità OLYMPUS TJF-Q180V DUODENOSCOPE per la riparazione Fabbrica

OLYMPUS TJF-Q180V DUODENOSCOPE per la riparazione

OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and ...
Qualità Colonoscopio Olympus CF-240L per la riparazione Fabbrica

Colonoscopio Olympus CF-240L per la riparazione

Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage...
Qualità Endoscopio Karl Storz 27005BA per riparazione Fabbrica

Endoscopio Karl Storz 27005BA per riparazione

Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Qualità Toshiba PST-25BT Probe ad Array per la riparazione Fabbrica

Toshiba PST-25BT Probe ad Array per la riparazione

Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common ...
Qualità Testa telecamera STRYKER 1188-210-105 per riparazione Fabbrica

Testa telecamera STRYKER 1188-210-105 per riparazione

STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and ...
Qualità Capocamera STORZ H3-ZA da riparare Fabbrica

Capocamera STORZ H3-ZA da riparare

STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3...
Qualità Sonda endocavitaria GE RIC6-12-D per riparazione Fabbrica

Sonda endocavitaria GE RIC6-12-D per riparazione

GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts ...
Qualità Sonda ecografica Hitachi C41V1 per riparazione Fabbrica

Sonda ecografica Hitachi C41V1 per riparazione

Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens ...
Qualità Sonoscape 8-L743 sonda lineare da riparare Fabbrica

Sonoscape 8-L743 sonda lineare da riparare

Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage ...
Qualità Sonda addominale Sonoscape C362 per riparazione Fabbrica

Sonda addominale Sonoscape C362 per riparazione

Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens ...
Qualità Sonda addominale GE 4C-RS da riparare Fabbrica

Sonda addominale GE 4C-RS da riparare

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...
Qualità Careutech BK Medical 9048 Sonda Ecografica Per Riparazione Fabbrica

Careutech BK Medical 9048 Sonda Ecografica Per Riparazione

BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe ...
Qualità Parti di apparecchiature mediche Sonda lineare Sonoscape L743 per riparazione Fabbrica

Parti di apparecchiature mediche Sonda lineare Sonoscape L743 per riparazione

Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common ...
Qualità Sonda ecografica robotica BK Medical 9096 per riparazione Fabbrica

Sonda ecografica robotica BK Medical 9096 per riparazione

BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution ...
Qualità Sonda Samsung Medison 3D4-9 SN P0N7  Per Riparazione Fabbrica

Sonda Samsung Medison 3D4-9 SN P0N7 Per Riparazione

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D...
Qualità Casse per la riparazione di dispositivi medici di GE RIC5-9-D Fabbrica

Casse per la riparazione di dispositivi medici di GE RIC5-9-D

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...
Qualità Filtro angolare e tubo elicoidale CF-HQ190L usurato 20% fascio di luce rotto Fabbrica

Filtro angolare e tubo elicoidale CF-HQ190L usurato 20% fascio di luce rotto

Products: CF-HQ190L For Repair Problem & Solutoin: CF-HQ190L - Light bundle 20% Broken Knob Leaked The angle wire and helix tube Worn Biospy channel Worn Rubber Cracked Light guide tube Indented Solution Replace Biopsy Channel, Wire, Light Guide Tube, Knob, Rubber Applications: When it comes to ...
Qualità Casse di riparazione di apparecchiature mediche di Olympus CF-H190L Colonoscopio da riparare Fabbrica

Casse di riparazione di apparecchiature mediche di Olympus CF-H190L Colonoscopio da riparare

Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working ...
Qualità Equipaggiamento medico Olympus riparato soluzione definitiva per i servizi medici Fabbrica

Equipaggiamento medico Olympus riparato soluzione definitiva per i servizi medici

Product Description: Product Summary: The Medical Device Repair Cases product offered is a comprehensive solution for repairing and maintaining Olympus medical devices. With a warranty of 90 days, customers can have peace of mind knowing that their repaired devices are covered for an extended period...
Qualità OLYMPUS CF-HQ190L Casse per la riparazione di dispositivi medici Casse per la riparazione di attrezzi Fabbrica

OLYMPUS CF-HQ190L Casse per la riparazione di dispositivi medici Casse per la riparazione di attrezzi

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...
Prev 1 2 3 Prossimo